Ametek PLA-PLW Programming Manual Manual de usuario Pagina 57

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 74
  • Tabla de contenidos
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 56
53
UTILITY COMMANDS
UTILity:RANGe
This command selects and queries the range control mode.
Command Syntax UTILity:RANGe <bool>
Parameters <bool> range control mode.
0 : manual mode,
1 : auto mode,
Query Syntax UTILity:RANGe?
Returned Parameters <NR1> integer representing range control mode
UTILity:MEASure:SENSe
This command sets or queries the measurement sense (Local / Remote ).
Note: When input voltage is greater than 30 V, this command will cause an execution error.
Command Syntax UTILity:MEASure:SENSe <bool>
Parameters 0 | 1 | LOCal(0) | REMote(1)
Query Syntax UTILity:MEASure:SENSe?
Returned Parameters <bool> 0=LOCal
1=REMote
UTILity:TR:MODE
UTILity:TRANsient:MODE
This command selects the operating mode of the transient generator as follows.
TRIGger The transient switch to trigger levels upon receipt of a trigger.
CONTinuous The transient generator puts out a continuous pulse stream.
TOGGle The transient generator toggles between two levels upon receipt of a
trigger.
PULSe The transient generator puts out a single pulse upon receipt of a trigger.
Command Syntax TR:MODe <NR1>
TRANsient:MODe <NR1>
Parameters <NR1> transient mode
0:TRIGger
1:CONTinuous
2:TOGGle
3:PULSe
4:STEP
5:AUTO
6:ONCE
Query Syntax TR:MODe?
TRANsient:MODe?
Returned Parameters <NR1> integer value representing transient mode