21
Channel-Specific Current Command/Query. This command specifies the value of the
programmed current level for the TRANsient input when the electronic load is in the CC Mode. When
the Transient Subsystem is on, the electronic load input current will sw itch (under control of the
Transient Subsystem) between the main level and TLEVel at a rate determined by the present value
of SLEW.
In order for the input current level to switch, TLEVel must be set to a value greater than the main level.
If TLEVel is set to a value below the main level, no error is generated bu
t s
witching will not occur until
the main level is subsequently below the value of TLEVel.
Command Syntax CURRent:TLEVel <NRf+>
ITR <NRf+>
Parameters <NRf+>: Current transient level value. Refer to model specifications for
range | MIN | MAX
Query Syntax CURRent:TLEVel?
CURRent:TLEVel? MIN CURRent:TLEVel? MAX
ITR
ITR? MIN ITR? MAX
Returned Parameters <NRf+> returns the transient current level.
If the electronic load is not in CC Mode, the level will still be set, even if it
is less than the presently programm
ed input level.
[SOURce:]CURRent:[TRANsient:]TWIDth
This command sets the pulse width of the transients when the generator is in PULSe mode.
Note: This command will also change the CURR:FREQ value.
Command Syntax CURRent:[TRANsient:]TWIDth <NRf+>
Parameters Refer to model’s specification | MAX | MIN
Unit ms
Query Syntax CURRent:TRANsient:TWIDth?
CURRent:TWIDth?
Returned Parameters <NRf+>, pulse width time in ms.