16
calibration constants are lost unless they have been stored with CALibrate:SAVE.
Command Syntax CALibrate:STATe <bool>
Parameters 0 | 1 | OFF | ON
*RST Value OFF
Query Syntax CALibrate:STATe?
Returned Parameters <NR1>
CALibrate:TRANsient:OFFSet
This command can only be used in calibration mode. It is used to set the DAC count of
immediate level when enter transient calibration mode.
Command Syntax CALibrate:TRANsient:OFFSet <NR1>
Parameters <NR1> DAC count of immediate level
Query Syntax CALibrate:TRANsient:OFFSet?
Returned Parameters <NR1>
CALibrate:VALue:CURRent
This command is only used in calibration mode. It enters a calibration current that you obtain by
reading an external meter. You must first select a calibration level (with CALibrate:LEVel) for the value
being entered. These constants are not stored in nonvolatile memory until they are saved with
CALibrate:SAVE. If CALibrate:STATE OFF is programmed without a CALibrate:SAVE, the previous
calibration constants are restored.
Co
mmand Syntax CALibrate:VALue:CURRent <NRf>
Parameters <NRf> current value
CALibrate:VALue:VOLTage
This command is only used in calibration mode. It enters a calibration voltage that you obtain by
reading an external meter. You must first select a calibration level (with CALibrate:LEVel) for the value
being entered. These constants are not stored in nonvolatile memory until they are saved with
CALibrate:SAVE. If CALibrate:STATE OFF is programmed without a CALibrate:SAVE, the previous
calibration constants are restored.
Co
mmand Syntax CALibrate:VALue:VOLTage <NRf>
Parameters <NRf> voltage value