50
TRANSIENT COMMANDS
These commands program the transient generator of the electronic load. The transient
generator programs a second (transient) level at which the electronic load can operate
without changing the original programmed settings.
See also [SOURce:]CURRent:TLEVel, [SOURce:]RESistance:TLEVel, and
[SOURce:]VOLTage:TLEVel in the Input Commands section.
[SOURce:]TRANsient
This command turns the transient generator on or off.
Command Syntax TRANsient <bool>
Parameters 0 | 1 | OFF | ON
*RST Value OFF
Query Syntax TRANsient?
Returned Parameters <NR3>
[SOURce:]TR:MODe
[SOURce:]TRANsient:MODe
This command selects the operating mode of the transient generator as follows.
TRIGger The transient switch to trigger levels upon receipt of a trigger.
CONTinuous The transient generator puts out a continuous pulse stream.
TOGGle The transient generator toggles between two levels upon receipt of a
trigger.
PULSe The transient generator puts out a single pulse upon receipt of a trigger.
Command Syntax TR:MODe <NR1>
TRANsient:MODe <NR1>
Parameters <NR1> transient mode
0:TRIGger
1:CONTinuous
2:TOGGle
3:PULSe
4:STEP
5:AUTO
6:ONCE
Query
Syntax TR:MODe?
TRANsient:MODe?
Returned Parameters <NR1> integer value representing transient mode
Comentarios a estos manuales