DC Module Remote Programming ReFlex Power™ Programming Manual
M380056-03 Rev L 79
Source Trigger commands allow a voltage and/or current setpoint to be delayed for
later application. The trigger latency is on the order of 10-20 mSec. If multiple
voltage or current commands are received, the earlier ones of each type will be
overwritten, without error. These setpoints are not retriggerable.
Table 3-8. Software Triggered Setpoints
SOUR<n>:VOLT:TRIG <NRf>
Setup for a deferred SOUR<n>:VOLT <NRf> command
SOUR<n>:VOLT:TRIG?
Query returns pending Voltage setting, -0.0 if inactive
SOUR<n>:CURR:TRIG <NRf>
Setup for a deferred SOUR<n>: CURR <NRf> command
SOUR<n>:CURR:TRIG?
Query returns pending Current setting, -0.0 if inactive
TRIG<n>:TYPe <NRi>
0 = VOLT, 1 = CURR, 2 = Both, Apply the setpoints
3.13.2
V
OLTAGE AND
C
URRENT
S
OFTWARE
R
AMPS
C
OMMAND
S
UMMARY
SOURce<n>
:VOLTage
:RAMP <NRf>,<NRf>,<NRf>
:CURRent
:RAMP <NRf>,<NRf>,<NRf>
:VOLTCURR
:RAMP <NRf>,<NRf>,<NRf>,<NRf>
SOURce<n>
:VOLTage
:RAMP
:TRIGger <NRf>,<NRf>,<NRf>
:TRIGger?
:CURRent
:RAMP
:TRIGger <NRf>,<NRf>,<NRf>
:TRIGger?
:VOLTCURR
:RAMP
:TRIGger <NRf>,<NRf>,<NRf>,<NRf>
:TRIGger?
TRIGger<n>
:TYPe 3
:ABORt
Voltage and Current ramps can be setup and invoked on a software basis, these
commands use the same hardware that is used for LIST/TRIGGER functions, so
these functions cannot be used or intermixed at the same time. Issue a
LIST<n>:CLOSe command to reset the LIST/TRIGGER subsystem when usage of
these commands is done.