6.5.2 CALIBRATION
SCPI
C
OMMAND
R
EFERENCE
CALibrate<n>
CALIBRATE sub-commands
:UNLock <string>
Enables the EEPROM memory for storing calibration constants.
The enable string is <“6867”>.
:STORe
Stores the calibration constants in the EEPROM memory.
:LOCK
Disables access to the EEPROM memory, thereby preventing
storing of calibration constants.
:POSitive15 <0+NR1>
Calibration for Positive 15 DC supply. Value is between 0 to 255
:POSitive15?
Query the pot setting
:POSitive5 <0+NR1>
Calibration for Positive 5 DC supply. Value is between 0 to 255
:POSitive5?
Query the pot setting
:NEGative15 <0+NR1>
Calibration for Negative 15 DC supply. Value is between 0 to 255
:NEGative15?
Query the pot setting
:NEGative5 <0+NR1>
Calibration for Negative 5 DC supply. Value is between 0 to 255
:NEGative5?
Query the pot setting
:MODule
:LASTCAL <Date>
Set the last calibration date
:LASTCAL?
Query the last calibration date
:NEXTCAL <Date>
Set the next calibration date
:NEXTCAL?
Query the next calibration date.
1
Calibration is required to be run twice, once with Sense relays closed and with Sense
relays open, then Store the calibration constants.