ReFlex Power™ Programming Manual Controller Module Remote Programming
48 M380056-03 Rev M
&
&
&
&
&
&
7 6 5 4 3 2 1 0
7 6 5 4 3 2 1 0
&
&
&
5 Standard Event Status register
4,2 Error Event Queue / Message Available
1 Protection Event Flag
0,3,6,7 Reserved
4 Command execution error
0 *RST required after a cold start
Sticky bits, cleared on Query
='s 1, if any bit is set
7 ----
Protection Event Register
Protection Event Enable Register
Enable Standard Event register
standard Event Status Register
ST
atus
B
Service Request Enable / STatus Byte register bits
Returns a 96 bit hexnumber, with
a '1' bit set for each module
requesting an SRQ (96..1)
Returns a non-sticky version
of the PROT:EVENT Register
3 Output Voltage Fault
1 Constant Current Operation
0 Constant Voltage Operation
7 Over Current Fault
6 Volt|Curr Mode Fault
5 Module Disabled
Note 1: Any changes in bits 7-3, 1-
Note 2: Short duration mode changes of < 50 mSec. may
not be
detected. For the Load Module: Bit 0 is for
Resistance mode operation.
Returns a 96 bit hex number, with
a '1' bit set for each module with
non-zero fault register (96..1)