AC Module Remote Programming ReFlex Power™ Programming Manual
M380056-03 Rev L 109
Set a different voltage for each member of the phase group:
SOUR12:VOLT 110 // Set AC module in slot 12 to 110 volts
SOUR8:VOLT 90 // Set AC module in slot 8 to 90 volts
*OPC12? // Send next command after a ‘0’ is returned
*OPC8? // Send next command after a ‘0’ is returned
SOUR4:VOLT 75 // Set AC master module in slot 4 to 75 volts and
execute synchronous voltage change on all three
units
*OPC4? // Send next command after a ‘1’ is returned
… // Place other commands here
SYST:GROUP:DELETE 1004 // Delete 1004 group
or
SYST:GROUP:DELETE:ALL // Delete all groups
4.12.2 E
XAMPLE CREATING A
P
ARALLEL
G
ROUP
Create a Parallel group using three AC modules (units at slot 4, 8,12)
SYSTem:GROup:DEFine:PARallel 4,8,12 // Create the parallel group
SYSTem:GROup:CAT:PAR? // Query the group
OUTP1004:STATE 1 // Turn on all the units of the group
*OPC1004? // Send next command after a ‘1’ is returned
SOUR1004:VOLT 120 // Set the group to 120 volts
*OPC1004? // Send next command after a ‘1’ is returned
SOUR1004:CURR 20 // Each module may supply a maximum of 7.14 Amps. In
parallel group, the maximum current for the group is the
number of group members * the maximum current. (Refer
to specifications for high range and low range current.
*OPC1004? // Send next command after a ‘1’ is returned
… // Place other commands here
SYST:GROUP:DELETE 1004 // Delete 1004 group
or
SYST:GROUP:DELETE:ALL // Delete all groups