ReFlex Power™ Programming Manual DC Module Remote Programming
72 M380056-03 Rev M
Figure 3-6 and Table 3-6 specify the trigger out signals.
Figure 3-6. Output Trigger Timing
Table 3-6. Output Trigger Specifications
Trigger
Range
2
(Volts)
Low Level
3
(Volts)
High Level
(Volts)
Timing
(Microseconds)
Output Voltage and Current
1
+5.5 V to –0.7V
Vol <= 0.3V @
1mA
Voh >= 4.13
@ 1mA
Trigger Latency
4
Tlout 5
Positive Assertion Period
Touth
2
Negative Assertion Period
Toutl 2
Trigger Recovery Time
Trt > Toutx + 4
1
Typical high-level leakage current is 0.01 mA @ 5.5V DC.
2
All voltages are with respect to the associated return signal.
3
Maximum low-level sink current is 7 mA.
4
Trigger latency is the maximum time from selected event to when the trigger is issued.