4.16 Sense Subsystem - Sweep [3Pi controller only]
This subsystem controls the measurement current range, the data acquire sequence, and
the harmonic measurement window of the AC source.
Subsystem Syntax
SENSe
:SWEep
:OFFSet <n> Define trigger points relative to the start of
the digitizer data record
:TINTerval? Query the digitizer sample spacing
SENSe:SWEep:OFFSet
This command defines the trigger point relative to the start of the returned data record when
an acquire trigger is used. The values can range from:
MX Series I: -104 msec to 1000 msec in a single-phase configuration and from -312
msec to 1000 msec in a three-phase configuration.
MX Series II and RS Series: -42 msec to 1000 msec in a single-phase configuration and
from -128 msec to 1000 msec in a three-phase configuration.
When the value specified is negative (less than 0 msec), the values in the beginning of the
data record represent samples taken prior to the actual trigger moment.
Command Syntax SENSe:SWEep:OFFSet <NRf+>
Parameters MX Series I: -104 to 1000 for single phase configuration
-312 to 1000 for three phase configuration
MX/RS Series II:-42 to 1000 for single phase configuration
-128 to 1000 for three phase configuration
*RST Value 0
Examples SENS:SWE:OFFS -5
Query Syntax SENSe:SWEep:OFFSet?
Returned Parameters <NR2>
Related Commands SENS:SWE:TINT? MEAS:ARR