6.6.1 Voltage and Current Measurements
To the rms voltage or current, use:
MEASure:VOLTage?or
MEASure:CURRent?
To measure the dc voltage or current, use:
MEASure:VOLTage:DC?or
MEASure:CURRent:DC?
To measure the maximum current amplitude and the current crest factor, use:
MEASure:CURRent:AMPLitude:MAXimum?
MEASure:CURRent:CREStfactor?
To reset the peak measuremnt , use
MEAS:CURR:AMPL:RESet
6.6.2
Power Measurements
The MEASure queries can return real, and apparent power measurements as well as dc
power and power factor using the following commands:
MEASure:POWer:AC:APParent? measures the AC component of apparent
power in VA
MEASure:POWer:AC:REAL? measures the in-phase component of
power in watts
MEASure:POWer:AC:PFACtor? returns the output power factor
MEASure:POWer:DC? measures the dc component of power
6.6.3
Frequency Measurements
To measure the frequency, use the following command:
MEASure:FREQuency?