Ametek BPS Series Programming Manual Manual de usuario Pagina 82

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 226
  • Tabla de contenidos
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 81
AMETEK Programmable Power BPS / MX / RS Series SCPI Programming Manual
82 Manual P/N 7003-961 Rev AA
FREQuency:SLEW
This command sets the rate at which frequency changes for all programmed changes in
output frequency. Instantaneous frequency changes can be obtained by sending
MAXimum.
Command Syntax [SOURce:]FREQuency:SLEW[:IMMediate] <NRf+>
Parameters 0.01 to 1E9MAXimum
Unit Hz (Hertz per second)
*RST Value MAXimum
Examples FREQ:SLEW:IMM 75FREQ:SLEW MAX
Query Syntax [SOURce:]FREQuency:SLEW?
Returned Parameters <NRf>
Related Commands FREQ:SLEW:MODE FREQ
FREQuency:SLEW:MODE
This command determines how the frequency slew rate is controlled during a triggered
output transient. The choices are:
FIXed The frequency slew rate is unaffected by a triggered output transient.
STEP The frequency slew rate is programmed to the value set by
FREQuency:TRIGgered when a triggered transient occurs.
PULSe The frequency slew rate is changed to the value set by
FREQuency:TRIGgered for a duration determined by the pulse commands.
LIST The frequency slew rate is controlled by the frequency list when a triggered
transient occurs.
Command Syntax [SOURce:]FREQuency:SLEW:MODE<mode>
Parameters FIXed | STEP | PULSe | LIST
*RST Value FIXed
Examples FREQ:SLEW:MODE FIX
Query Syntax [SOURce:]FREQuency:SLEW:MODE?
Returned Parameters <CRD>
Related Commands FREQ FREQ:SLEW:TRIG
Vista de pagina 81
1 2 ... 77 78 79 80 81 82 83 84 85 86 87 ... 225 226

Comentarios a estos manuales

Sin comentarios