DLM-E Series and DCS-E Series Isolated Analog Programmer Operation
2.3.5 Resistance Programming of Output Current
SW1 Switch Settings
SW1-2 ON (1) = 0–5K
2101213 1
25 22
14
Set up for resistance programming of the output current as follows:
1. Set SW1-2, I PROG, to ON (1) for 0-5VDC programming range.
2. Connect the external programming resistance, 0-5k, to the ISOLATED ANALOG
PROGRAMMER connector, J4, from Pin-10 to Pin-12.
3. Connect a jumper from Pin-22 to Pin-10 to connect the 1mA current source.
4. Connect Pin-1, ANALOG-CONTROL, of the ISOLATED ANALOG PROGRAMMER
connector, J4, to Pin-2 to enable remote control.
5. Program the other parameters to the desired limit values.
M51A Option 2-11