
Isolated Analog Programmer Operation DLM-E Series and DCS-E Series
2.3.2 Voltage Source Programming of Output Current
SW1 Switch Settings
SW1-2 ON (1) = 0–5V
SW1-2 OFF (0)= 0–10V
2101213 1
25
14
Set up for voltage source programming of the output voltage as follows:
1. Set SW1-2, I PROG, to ON (1) for 0-5VDC programming range.
2. Set SW1-1, V PROG, to OFF (0) for 0-10VDC programming range.
3. Connect the external programming voltage source to the ISOLATED ANALOG
PROGRAMMER connector, J4, with positive to Pin-10 and negative to Pin-12.
4. Connect Pin-1, ANALOG-CONTROL, of the ISOLATED ANALOG PROGRAMMER
connector, J4, to Pin-2 to enable remote control.
5. Program the other parameters to the desired limit values.
2-8 M51A Option
Comentarios a estos manuales