User Manual – Rev H California Instruments
46 Avionics Software
3.4.2 DO160 Section 16, Rev F:
3.4.2.1 Table A(CF)115:
Normal Operating Conditions (ac)
1 Phs Voltage and Frequency (ac)
100V to 122V, 390Hz to 410Hz
Apply for 30 mins each test
3 Phs Voltage and Frequency (ac)
100V to 122V, 390Hz to 410Hz
Apply for 30 mins each test
Emergency Operating Conditions (ac)
1 Phs Voltage and Frequency (ac)
100V to 122V, 360Hz to 440Hz
Apply for 30 mins each test
3 Phs Voltage and Frequency (ac)
100V to 122V, 360Hz to 440Hz
Apply for 30 mins each test
Normal Operating Conditions (ac)
Vamp = 0.18V to 1.24V, Mrate = 1Hz to 200Hz
Frequency Modulation (ac)
Famp = 4.9 Hz to 0.2Hz, Mrate = 0.014Hz to 100Hz
Momentary Power Interruptions (ac)
Equipment with Digital Circuits
Momentary Power Interruptions (ac)
16.5.1.5.1 - Normal Voltage Surge (ac)
16.5.1.5.2 - Normal Frequency Surge (ac)
16.5.1.7 - 1 Phs Voltage DC Content (ac)
Apply for 30 mins each test
16.5.1.7 - 3 Phs Voltage DC Content (ac)
Apply for 30 mins each test
8% +2%/-0% VTHD - Clipped Sine, 400Hz
Apply for 30 mins each test
Abnormal Operating Conditions (ac)
Apply for 5 mins each test
Apply for 5 mins each test
1 Phs Voltage and Frequency (ac)
100V to 122V, 370Hz to 430Hz
Apply for 5 mins each test
3 Phs Voltage and Frequency (ac)
100V to 122V, 370Hz to 430Hz
Apply for 5 mins each test
Momentary Undervoltage (ac)
Apply for 7 seconds each test
16.5.2.3.1 - Abnormal Voltage surge
Figure 16.5 - apply at 10 sec
intervals
16.5.2.3.1 - Abnormal Frequency transients
0V to 115V, 320Hz to 480Hz
Figure 16.5 - apply at 10 sec
intervals