Ametek SFA Series Ethernet Programming Manual de usuario Pagina 150

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 159
  • Tabla de contenidos
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 149
Calibration SG Series Programming
7-16
M550129-03 Rev K
7. Set the programming source to 0.5V ±1mV.
8. Verify that the unit is set to 10% ± 0.25% of full-scale output voltage. If
necessary adjust R35 (on the isolated interface adapter) for 10% output voltage.
9. Repeat the steps above as required to obtain the required accuracy.
10. Remove all connections to the remote analog control connector, J1 (Figure 7-6).
Place the jumper between pins 5 and 6. This reverts the SGI series unit back to
local, front panel control.
Figure 7-6. Analog Control Connector (J1)
Vista de pagina 149
1 2 ... 145 146 147 148 149 150 151 152 153 154 155 ... 158 159

Comentarios a estos manuales

Sin comentarios