Vista de pagina 146
SG Series Programming Calibration
M550129-03 Rev K
7-13
Preparation:
1. Remove prime AC power.
2. Connect the power supply to a precision current shunt (Figure 7-2) capable of
a minimum of 10% above full rated output. A fan may be required to cool the
shunt.
J3
SHUNT
SUPPLY
OUTPUT
TERMINALS
+
-
J3-1: SENSE (+)
J3-2: SENSE (-)
J3-3: SENSE (-), 800V ONLY
Figure 7-2. Precision Current Shunt