Programming Manual Lx \ Ls Series II
193
8.4.4 DO160 Abnormal Test Commands
DO160:ABNormal:VOLTage:MINimum
This command will set the voltage to the minimum level for the abnormal operation.
Command Syntax DO160:ABNormal:VOLTage:MINimum
Parameters none
Examples DO160:ABN:VOLT:MIN
Query Syntax none
DO160:ABNormal:VOLTage:MAXimum
This command will set the voltage to the maximum level for the abnormal operation.
Command Syntax DO160:ABNormal:VOLTage:MAXimum
Parameters none
Examples DO160:ABN:VOLT:MAX
Query Syntax none
DO160:ABNormal:VOLTage:UNDer
This command will set the voltage to the under voltage level for the abnormal
operation.
Command Syntax DO160:ABNormal:VOLTage:UNDer
Parameters none
Examples DO160:ABN:VOLT:UND
Query Syntax none
DO160:ABNormal:VOLTage:OVER
This command will set the voltage to the surge level for the abnormal operation.
Command Syntax DO160:ABNormal:VOLTage:OVER
Parameters none
Examples DO160:ABN:VOLT:OVER
Query Syntax none
DO160:ABNormal:VOLTage:UNBalance
This test will change the output voltage for phase A and B to 134V and phase C to a
voltage lower by a value specified by an offset. This command applies only to test
standard selection RTCA2. Refer to the User Manual for details.
Command Syntax DO160:ABNormal:VOLTage:UNBalance
Parameters none
Examples DO160:ABN:VOLT:UNB
Query Syntax none